GE
D
IRECTION
5535208-100, R
EV
. 2
LOGIQ E9 S
ERVICE
M
ANUAL
Chapter 8 Replacement procedures
8 - 111
8-6-3-3
LCD Monitor V2 Arm assembly replacement - R5.x and later
Table 8-69 LCD Monitor V2 Arm assembly removal - R5 and later
Steps
Corresponding Graphic
1.
Remove the LCD Monitor.
2.
WARNING
WARNING
WHEN REMOVING THE LCD ARM, KEEP
THE ARM LOCKED, AS SHOWN TO
ENSURE THE LCD ARM ASSEMBLY IS
IN THE LOCKED POSITION UNTILL THE
ARM AND LCD ARE INSTALLED. THE
SPRINGS TO SUPPORT THE LCD CAN
CAUSE THE ARM TO SPRING OPEN
CAUSING SEVERE PERSONAL INJURY
AND PROPERTY DAMAGE.
3.
Remove the Bulkhead Cover.
4.
Disconnect the LCD and Arm cables; Video
(
P10
) and power (
P3
) from the LCD and the
and ground Cable from the Arm.
The Video Cable has a retainer and screw.
Loosen the screw with a small, flat blade
screwdriver. The ground Cable uses a
Phillips screw.
Содержание 5205000
Страница 1: ...10 20 14 GEHC_FRNT_CVR FM LOGIQ E9 SERVICE MANUAL Part Number 5535208 100 Revision Rev 2 ...
Страница 2: ......
Страница 21: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 19 ZH CN KO ...
Страница 198: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 4 54 Section 4 8 Site Log This page was intentionally left blank ...
Страница 807: ......
Страница 808: ......