Section 3-9 - Connectivity Setup and Tips
7.
In the
Dataflow
tab, add your Storage Commitment service to the same dataflow that handles
your Image Storage service.
Storage Commitment includes “reverse negotiation” with the receiving end – the Image
Storage service. Basically, the receiver needs to know the AE Title, IP address and Port
Number of the . Refer to the documentation of the Image Storage service for setting up Storage
Commitment.
NOTE: The Store Commit and its Associated Storage (A) MUST BE in the same Dataflow!
DICOM Job Spooler - STC identified
To confirm a successful Storage Commitment, press
F4
to inspect the spooler after you send images
to the Image Storage service. An entry of:
•
Type: STC
•
Contents: Value equals the number of images stored, plus one if a Structured Report was sent.
•
Status: “Done” or “Success”
•
Error: no indications
confirms a successful storage commitment.
Table 3-23 DICOM Storage Commitment Service Setup and Use