Chapter 10 Care & Maintenance
10 - 25
10-6-10-1 Generic Procedure on Probe Leakage Current (cont’d)
Follow the test conditions and test limits described in:
•
Table 10-9 on page 10-11
•
Table 10-10 on page 10-11
as applicable for every probe.
Keep a record of the results with other hard copies of maintenance data using
Table 10-15 on page 10-25
Table 10-15 on page 10-25
shows a typical format for recording ultrasound probe source leakage
current.
•
Table 10-9 on page 10-11
•
Table 10-10 on page 10-11
as applicable.
NOTE:
Values in italics font are given as examples only.
Table 10-15 Typical data format for recording probe (source) leakage
CAUTION
Equipment damage possibility. Never switch the Polarity or the status of the Neutral when the
Ultrasound system is powered on.Power off the Ultrasound system, allow the stored energy to
bleed down, and turn the circuit breaker off BEFORE switching the POLARITY switch and/or the
NEUTRAL switch on the leakage meter to avoid possible power supply damage
Unit under test____________________________________ Date of test:_____________________
Test Conditions
Probe as measured in saline bath
System
Power Grounding/PE
4C
i12L
TS
E8C
Off
Closed
Off
Open
On
Closed
On
Open
Содержание 5205000
Страница 1: ...10 20 14 GEHC_FRNT_CVR FM LOGIQ E9 SERVICE MANUAL Part Number 5535208 100 Revision Rev 2 ...
Страница 2: ......
Страница 21: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 19 ZH CN KO ...
Страница 198: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 4 54 Section 4 8 Site Log This page was intentionally left blank ...
Страница 807: ......
Страница 808: ......