Section 8-8 - Casters and Brakes replacement
8-8-3-4
Front Casters removal
1.) Turn Front Casters so they are pointing forwards.
2.) Activate Direction Lock.
3.) Put the Bevel Edged Board on the floor.
4.) Pull the system backwards up the board incline. This will lift the Rear Wheel on the opposite side of
the System from the floor.
5.) Turn the Rear Caster that stands on the Bevel Edged Board in the direction as shown in
Figure 8-65
.
6.) The system is now nearly balanced between one Front and one Rear Caster.
7.) Make the LOGIQ E9 rest on both Rear Casters and lift the Front Caster.
8.) Put the Wooden Wedge under the chassis. This will stabilize the LOGIQ E9 with one Front Caster
free from the floor.
9.) Unscrew and remove the fixing screws (T-45) for the Front Caster that is free from the floor. Save
the screw for later use.
10.)Remove the Caster.
11.)Disconnect the lock by unscrewing the hex nut that holds it to the rod.
WARNING
WARNING
Prior to elevating scanner, verify that the floating Operator Panel is locked in its lowest,
parking position.
WARNING
WARNING
Use extreme caution as long as LOGIQ E9 is un-stable, not resting on all four Casters.
Figure 8-66 Pull system backwards up the board incline
Содержание 5205000
Страница 1: ...10 20 14 GEHC_FRNT_CVR FM LOGIQ E9 SERVICE MANUAL Part Number 5535208 100 Revision Rev 2 ...
Страница 2: ......
Страница 21: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 19 ZH CN KO ...
Страница 198: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 4 54 Section 4 8 Site Log This page was intentionally left blank ...
Страница 807: ......
Страница 808: ......