GE
D
IRECTION
5535208-100, R
EV
. 2
LOGIQ E9 S
ERVICE
M
ANUAL
8 - 124
Section 8-6 - Replacing Top Console Parts
8-6-5
WLAN replacement - R2.x and R3.x only
Table 8-77 Manpower / Time and Tools
Manpower /
Time
Tools
One person /
15 minutes
Refer to:
8-2-5 "Tools needed for servicing the LOGIQ E9" on page 8-5
• Appropriate PPE
• ESD
Table 8-78 Preparations and Preparation Link
Preparations - you must perform the following steps
NOTICE
ENERGY CONTROL AND POWER LOCKOUT FOR LOGIQ E9
WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS EXPOSURE TO VOLTAGE
GREATER THAN 30 VOLTS:
1. TURN OFF THE SCANNER.
2. UNPLUG THE SYSTEM.
3. MAINTAIN CONTROL OF THE SYSTEM POWER PLUG.
4. WAIT FOR AT LEAST 20 SECONDS FOR CAPACITORS TO DISCHARGE AS THERE ARE
NO TEST POINTS TO VERIFY ISOLATION. THE AMBER LIGHT ON THE OP PANEL ON/OFF
BUTTON WILL TURN OFF.
5.
DISCONNECT THE EPS (EXTENDED POWER SHUTDOWN) BATTERY AT J3, OR THE
BATTERY FROM THE CHARGEBOARD AT PCN1 WHEN WORKING IN THE BEP. THIS
SHOULD BE DONE WHENEVER THE BEP IS OPEN AND THE EPS OR CHARGEBOARD
IS EXPOSED AND CHANGING PARTS.
Beware that the Main Power Supply, Extended Power Shutdown or ChargeBoard and BEP
may be energized even if the power is turned OFF if the cord is still plugged into the AC
Outlet.
1.Select
Utility
->
Connectivity
->
TCP/IP
->
Wireless Network
->
Configuration
2.As a precautionary step:
a.Document all the information regarding configuration of the WLAN.
b.Select the
Properties
tab.
c.Record the following information (provided by the Customer site IT):
i. Network Name (SSID)
ii. Network Authentication
iii.Data Encryption
iv.Network Key
v.Key Index
vi.Specify Yes/No to “The Key is Provided Automatically”
vii.IEEE 802.1x Authentication information
viii.EAP type
d. Press Close.
3. Power down the system.
4. Move the User Interface (Top Console) to the lowest position.
5. Disconnect the mains power cable from the wall outlet and all Probes and External I/O Cabling.
Preparation Link (if you need more information):
4-2-3 "Power shut down" on page 4-7
.
Signed
Date
TAG
&
LOCKOUT
Содержание 5205000
Страница 1: ...10 20 14 GEHC_FRNT_CVR FM LOGIQ E9 SERVICE MANUAL Part Number 5535208 100 Revision Rev 2 ...
Страница 2: ......
Страница 21: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 19 ZH CN KO ...
Страница 198: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 4 54 Section 4 8 Site Log This page was intentionally left blank ...
Страница 807: ......
Страница 808: ......