Chapter 3 LOGIQ E9 Setup
3 - 77
Encoded Value in the “Status” tag
Other times, however, the actual reason for an issue is often encoded in innocuous-looking entries
near the end of the exchange. For example, the above situation involved the attempting to send an
image to Viewpoint directly via a print button, without a patient id. Viewpoint’s “C-STORE-RSP”
message to the indicated, by an encoded value in the “Status” tag, that it did not store the image.
(continued on next page)
Table 3-30 Network “Sniffer”
Содержание 5205000