GE
D
IRECTION
5535208-100, R
EV
. 2
LOGIQ E9 S
ERVICE
M
ANUAL
Chapter 5 Components and Functions (Theory)
5 - 23
Section 5-7
Power distribution
5-7-1
Purpose of this section
The power distribution within the LOGIQ E9 is described in this section.
5-7-2
Main Power Supply
5-7-2-1
General description
The Main Power Supply’s main task is to galvanically isolate the scanner from the on-site Mains Power
System and to supply the various internal subsystems with AC or DC power.
Figure 5-12 Main Power Supply
MLP
Line Filter
Rectifier
Inrush Current
Limier
PFC
Stand By
+ 5Vstb (12W)
+ 24V
24V
± 48V
+ 48V
+ 15V
+ 12V
+ 6V
TLP
± 0 – 95V
± 0 – 95V
2X20W
IEC60601-1
Isolation
400V
90-250Vac
0-140V
0-140V
130W
Inverter
110V
120VA
113 – 400 V
+ 48V
– 15V
– 6V
- 48V
PMX
± 100 V
8W
+/- 48V
SELV outputs (<60Vdc)