7-5-6-8
Touch Screen Tab
Test the Touch Screen functionality by pressing on the Touch Screen of the upper operator panel. Verify
that the press is indicated in the black window by a highlighted circle. See:
Figure 7-45
or
Figure 7-46
.
Figure 7-45 Touch Screen Tab - R4 and earlier
Figure 7-46 Touch Screen Tab - R5 and later
Check calibration
:
Touch screen and verify dots follow/
Calibrate: Press and hold each touch
target until it moves to next target.
Select Touch screen test, press each
target as they appear/Calibrate: Press
Run Windows calibration. Press very
lightly on each target until it moves to
next location.