Operating Manual
BASIC SETTINGS
Software 2.6
B|BRAUN
Chapter 7
Page 3 of 4
SHARING EXPERTISE
2.6_CA-EN_03
Def. Ratio Plasma/Blood
This parameter sets the percentage share of
plasma flow to blood flow during the separation of
plasma. The setting is performed in steps of 1% in
the range from 10% to 40%. The default setting is
20 %.
The plasma/blood ratio is relevant to patient
safety, therefore confirmation of its change is
required.
Def. PPL Threshold
This parameter sets the limiting value for the
automatic plasma flow adaptation during therapy.
The setting is performed in steps of 5 mmHg in the
range from -20 to 120 mmHg. The default setting
is 20 mmHg.
Def. Ratio Dialysate/Plasma:
This parameter sets the ratio of the dialysate flow
in relationship to the plasma flow during the
therapy and reinfusion. The setting is performed in
steps of 1 in the range from 4 to 12. The default
setting is 4.
The ratio of dialysate/plasma is a parameter
relevant to patient safety, therefore confirmation
of its change is required.
Def. Plasma Reinfusion Flow
This parameter sets the Plasma Reinfusion Flow
default value on the Default screen: in the range
of 10-50 ml/min (First default: 30 ml/min). In
every therapy the Reinfusion Flow is set to this
default value after a new therapy selection.
Date
Date, month and year are set successively.
Time
Hours and minutes are set successively.
The modification of the following parameters must be
confirmed with the
key since they are relevant
to patient safety:
Def. Ratio Plasma/Blood
Def. Ratio Dialysate/Plasma.
OK