Operating Manual
PREPARATION
Software 2.6
B|BRAUN
Chapter 3
Page 2 of 9
SHARING EXPERTISE
2.6_CA-EN_03
3.1 SWITCHING ON AND SELF-TEST
Switching on
Switch on the H.E.L.P. Plasmat® Futura with the
On/Off switch on the rear of the machine.
Make sure that the machine brake is locked during
the treatment.
Hardware Self-Tests
After the machine has been switched on, the system
performs a series of hardware self-tests.
The screen shows the
controller tests
on the left side
and the
supervisor tests
on the right side.
The <
Retest
> menu item blinks during the self-test.
Positive self-test:
•
All tested positions are marked with "PASSED".
•
All three rows of numbers are completely
presented in the correct sequence (0 1 2 3 4 5 6 7
8 9) completely and in the three fonts that can be
displayed by the machine.
After a positive self-test, the <
End
> menu item is
automatically activated. Confirm with the
key
to change to the Start screen
Negative self-test:
•
The affected positions are marked "Failed" and/or
•
The rows of numbers are not in the correct
sequence or incomplete.
The <
Retest
> function is automatically selected in
this case. Confirm with the
key to start the retest.
See Annex 4 for detailed information concerning the
self-tests.
The different acoustic signals of the machine are also
tested during the self-test. Please make sure the
acoustic signals are audible.
Make sure that all LEDs are blinking.
RETEST