9.3 TECHNICAL DATA
9.3.1 Technical Data - General
Machine dimensions
Height:
1330
mm
Width:
500
mm
Weight:
520
mm
Weight
55
kg
Electrical connection
Rated voltage
110 – 240
V AC
Rated frequency
50/60
Hz
FI circuit-breaker
30
mA
Class of protection 1, type B, IP 21
The rated voltage must be identical with the
voltage specified on the type plate (e.g. 230
V AC, 50/60 Hz)
Power input
Rated current:
3.5
A max.
Classification
Type IIb according to Directive 93/42 EEC
Leakage currents
Ground leakage current:
< 500
µA
Patient leakage current:
< 100
µA
The allowed leakage currents may increase
when several machines are connected.
Operating conditions
Operating temperature;
+15 ÷ +35
o
C
Rel. humidity:
30 – 90
%
Atmospheric pressure:
700 – 1060
mbar
Storage conditions
Storage temperature;
- 20 ÷ +55
o
C
Rel. humidity:
10 – 90
%
Atmospheric pressure:
700 – 1060
mbar
Potential equalization
Connection according to DIN 42801 (EN 60-601/1)
Interface
RS 485 interface for the connection of an external PC by the
technical service or for therapy data collection and/or
monitoring (option, information on request)
The external PC must comply with the ICE
950 standard (or equivalent standards/
directives).
Electromagnetic compatibility
According to EN 60601-1-2 (IEC 601-1-2)
Housing material
Corrosion-resistant aluminium
Plastics ( polyurethane Baydur )