Operating Manual
PREPARATION
Software 2.6
B|BRAUN
Chapter 3
Page 4 of 9
SHARING EXPERTISE
2.6_CA-EN_03
3.2 PREPARING THE SOLUTIONS
H.E.L.P. 0.9% Nacl, physiological saline solution
Remove the outer packaging of the saline bag.
Fill a syringe with 1.5 ml heparin
(5.000 IU / ml).
Remove the cannula from the syringe.
Remove the screw cap from one of the Luer-lock
connectors of the bag and insert the syringe.
Break the seal of the bag.
Inject the 1.5 ml heparin into the saline bag.
Carefully mix the heparin with the saline solution.
Prepare the second bag in the same manner.
Bicarbonate Solution H.E.L.P. BicEL
Remove the outer packing from the bag.
Place the bag on a firm base and press the smaller
chamber of the bag with both hands until the seal
seam between the two chambers is opened over its
full length.
Move the bag several times to and fro so that the
two solutions are well mixed.
Prepare the other bag accordingly.
Acetate Buffer Solution
Remove the outer packaging of the acetate buffer
bag.
Fill a syringe with 30 ml H.E.L.P. heparin sodium
solution for extracorporeal application.
Remove the cannula from the syringe.
Remove a Luer-lock connector from the acetate
buffer bag and insert the syringe.
Break the seal.
Inject the 30 ml H.E.L.P. heparin sodium solution
into the acetate bag
Carefully mix the H.E.L.P. heparin sodium solution
with the acetate buffer.