Operating Manual
ALARMS AND PROBLEM CORRECTION
Software 2.6
B|BRAUN
Chapter 8
Page 4 of 27
SHARING EXPERTISE
2.6_CA-EN_03
Code
Alarm Text
Alarm Cause
Corrective Action
A 08
Safety air detector
(SAD) test failed
Hardware problem
Acknowledge the alarm (twice). If the alarm
is repeated switch the machine off and on
again.
If the problem cannot be solved, stop the
treatment immediately taking care to visually
inspect for air bubbles in the venous return
line.
Then inform technical service.
A 09
Weight system test
failed
Hardware problem
Acknowledge the alarm (twice). If the alarm
is repeated switch the machine off and on
again.
If the problem cannot be solved, stop the
therapy immediately then close the
treatment with reinfusion.
Inform technical service.
A 10
User interface not
communicating
Hardware problem
Acknowledge the alarm (twice). If the alarm
is repeated switch the machine off and on
again to eliminate a possible transient
failure.
If the problem cannot be solved, stop the
treatment immediately and inform technical
service.
A 13
Arterial pressure (PA)
not zero
Consumables already or
still mounted
Remove all consumables from the machine.
A 14
Prefilter pressure (PBE)
not zero
Consumables already or
still mounted
Remove all consumables from the machine.
A 15
Venous pressure (PV)
not zero
Consumables already or
still mounted
Remove all consumables from the machine.
A 16
Load cell not empty or
load cell error
Consumables already or
still mounted
Remove all consumables from the machine.
A 17
Line in SAD not empty
or SAD error
Consumables already or
still mounted
Remove all consumables from the machine.
A 18
Chamber in PCLD not
empty or PCLD error
Consumables already or
still mounted
Remove all consumables from the machine.
A 19
Chamber in HCLD not
empty or HCLD error
Consumables already or
still mounted
Remove all consumables from the machine.
A 20
Line in DAD not empty
or DAD error
Consumables already or
still mounted
Remove all consumables from the machine.
A 21
Power relay test failed
Defective hardware
Switch the machine off and on again and
restart the machine.
A 22
Heater relay test failed
Defective hardware
Switch the machine off and on again and
restart the machine.
A 25
Check correct insertion
return line
In the priming and rinsing
phase, a test is performed
to determine whether the
pump segment of the
plasma/buffer pump is
correctly inserted. This
test failed.
Ensure that:
The pump segment is correctly inserted in
the plasma/ buffer pump.