When <
Additional Functions
> is selected, a submenu
with the following selections is opened:
Stop bolus – active only while the heparin
bolus is administered
Heparin bolus – active during the therapy
Balance reset – active only for improper
balancing > 200 g (for a more detailed
description, see Problem Correction).
The active menu items are shown in black labeling,
and the inactive items in gray labeling. The selected
active field has a green background.
Heparin bolus
To administer a heparin bolus during the therapy,
select the <
Heparin Bolus
> menu item and
confirm the input with the
key.
The Warning window is displayed with the
following message: <
W33: Heparin bolus?
>
Confirm the message with the
key if you
wish to administer the heparin bolus.
If you do not wish to administer the heparin
bolus, wait for the Warning window to disappear
after 5 seconds.
OK
Warning
W33: Heparin bolus ? 10 ml
Press ‘OK’ to proceed!