Operating Manual
Installation and Technical Data
Software 2.6
B|BRAUN
Chapter 9
Page 14 of 17
SHARING EXPERTISE
2.6_CA-EN_03
9.3.6 Dialysing Circuit
Dialysate pump (DP)
Peristaltic roller pump with motor switch-off when the pump
cover is open
Delivery rate:
10 ÷ 200
ml/min
Delivery rate tolerance:
< ± 10
%
Working pressure range:
-140 ÷ +500
mmHg
Protection system:
Pump status and rate is monitored via a rotation detector.
Alarm override:
Not possible during the therapy
Acoustic alarm interval:
120
s
Dialysate inlet pressure (PDI)
Electronically measured by a pressure sensor and digitally
displayed
Measurement range:
- 500 ÷ +500
mmHg
Allowed tolerance
± 10
mmHg
Working range:
+60 ÷ +80
mmHg
During Therapy:
Alarm limits:
-50 ÷ +450
mmHg
Protection system:
Sensor test during preparation phase.
Alarm override:
Not possible during the therapy
Acoustic alarm interval:
120
s
Air detector (DAD)
Ultrasonic sensor on the dialysate line behind the dialysate
pump
Sensitivity:
Air for 800 ms
Protection system:
Sensor test during preparation phase.
Alarm override:
40 s after alarm
Acoustic alarm interval:
120
s