Operating Manual
PREPARATION
Software 2.6
B|BRAUN
Chapter 3
Page 9 of 9
SHARING EXPERTISE
2.6_CA-EN_03
Setting Up the Dialysate Line
(1)
Insert the heating bag into the plate warmer.
(2)
Connect the blue inflow line to the dialyzer.
Make sure that the Hansen connector is firmly
seated.
Note: Connect red with red and blue with blue!
(3)
Place the blue inflow line into the dialysate air
detector (DAD).
(4)
Insert the pump segment of the dialysate line
into the dialysate pump.
(5)
Screw on the pressure transducer.
(6)
Connect the prepared dialysate bag to the
distributor of the dialysate line and break the
seal.
(7)
Insert the dialysate inlet line into the provided
holder of the load cell.
1
2
4
3
5
6
7