Operating Manual
PRIMING AND RINSING
Software-Version 2.6
B|BRAUN
Chapter 4
Page 15 of 19
SHARING EXPERTISE
2.6_CA-EN_03
Press the
key to activate the parameter to be
changed. The field is shown with a red background
and white labeling. Perform the desired change using
the rotary knob and confirm it with the
key.
The changing of the following parameters must be
confirmed with the
key since they are relevant
for safety:
Plasma flow rate
Plasma volume
Balance
Heparin flow rate
Heparin bolus
PV MIN window (mmHg)
PV MAX window (mmHg)
Ratio Dialysate/Plasma
If a parameter is relevant for safety, the currently set
value is shown in the Setting window above the
setting range. In addition, the LEDs
above the
key
blink.
WARNING
Risk to patient due to blood loss since increasing
the PV MIN window elevates the likelihood of an
unrecognised removal of the venous access.
Do not cover the venous access.
Keep the patient under continuous
surveillance.
To quit the screen for setting the parameters, press
the
key. The cursor changes back to the menu
bar of the Parameter Overview screen and the menu
item <Start Priming>
If you do not perform any settings for more than
15 seconds, the screen automatically changes back to
the previously set screen.
OK
OK
OK
Setting
Plasma flow
%
20
[ 10 : 40 ]
W16: Press 'OK' to return to menu selection !