Operating Manual
ALARMS AND PROBLEM CORRECTION
Software 2.6
B|BRAUN
Chapter 8
Page 3 of 27
SHARING EXPERTISE
2.6_CA-EN_03
8.1.2 List of Alarms
If an alarm cannot be corrected with the measures
described, if it occurs frequently and you cannot
determine its cause or if a machine defect exists,
please inform technical service.
Code
Alarm Text
Alarm Cause
Corrective Action
A 01
Supervisor system not
working properly
Hardware problem
Acknowledge the alarm (twice). If the alarm
is repeated switch the machine off and on
again to eliminate a possible transient
failure.
If the problem cannot be solved, close the
treatment immediately and inform technical
service.
A 02
Deviation between
controller and
supervisor state
Hardware problem
Acknowledge the alarm (twice). If it is not
possible switch the machine off and on
again.
If the problem cannot be solved, restart the
machine completely or close the treatment
immediately.
If the problem cannot be solved with a
machine restart inform technical service.
A 03
Deviation of arterial
pressure between
controller and
supervisor
Calibration or hardware
problems
Acknowledge the alarm (twice).
If the problem cannot be solved, inform
technical service.
A 04
Deviation of venous
pressure between
controller and
supervisor
Calibration or hardware
problems
Acknowledge the alarm (twice).
If the problem cannot be solved, inform
technical service.
A 05
Deviation of weight
fluid between controller
and supervisor
Calibration or hardware
problems
Acknowledge the alarm (twice).
If the problem cannot be solved, inform
technical service.
A 06
Deviation of
temperature between
controller and
supervisor
Calibration or hardware
problems
Acknowledge the alarm (twice).
If the problem cannot be solved, inform
technical service.
A 07
Blood leak detector
(BLD) test failed
Hardware problem
Acknowledge the alarm (twice). If the alarm
is repeated switch the machine off and on
again.
If the problem cannot be solved, stop the
treatment as soon as possible while visually
inspecting for a possible blood leak in the
plasma line.
Inform technical service.