Operating Manual
ANNEX 2 – EXPLANATION OF PRESSURES
Software 2.6
B|BRAUN
Annex 2
Page 4 of 4
SHARING EXPERTISE
2.6_CA-EN_03
Pressure Limits
The following table shows the limiting value ranges of the pressures depending on the current phase of the
system.
Pressure
(mmHg)
Priming
Therapy BP RUN
Therapy BP STOP
Reinfusion BP RUN
Reinfusion BP STOP
Def
Def /
Auto
Min
Max
Def
Min
Max
Def /
Auto
Min
Max
Def
Min
Max
PA min
-150
-150
-350
80
-150
-350
80
-150
-350
80
-150
-350
80
PA max
190
100
0
200
200
-
-
100
0
200
200
-
-
PBE min
(2)
-450
PBE ref–40
-100
210
-100
-
-
PBE ref–40
(plasma reinf.)
PBE ref–60
(blood reinf.)
-100
210
-100
-
-
PBE max
(2)
250
PBE ref+80
-20
250
250
-
-
PBE ref+80
-20
250
250
-
-
PV min
(3)
-450
PV ref -
MinW
-10/0
250 -
MinW
-20
-
-
-20
-
-
-20
-
-
PV max
(3)
250
PV ref +
MaxW
PVmin +
MaxW
250
250
-
-
PV ref +
MaxW
PVmin +
MaxW
250
250
-
-
Pressure
(mmHg)
Priming
Therapy fluid side RUN
Therapy fluid side
STOP
Reinfusion fluid side RUN
Reinf. fluid side STOP
Def
Def
Min
Max
Def
Min
Max
Def
Min
Max
Def
Min
Max
TMP max
200
100
20
200
100
20
200
150
20
200
150
20
200
PPL min
-20
-10
-20
10
-100
-
-
-100
-
-
-100
-
-
PPL max
450
200
-
-
200
-
-
200
-
-
200
-
-
PPF min
-50
-20
-50
50
-250
-
-
-20
-50
50
-250
-
-
PPF max
450
450
-
-
450
-
-
450
-
-
450
-
-
PDF min
-50
-50
-50
0
-50
-
-
-50
-50
0
-50
-
-
PDF max
400
350
10
450
450
-
-
400
10
450
450
-
-
PDPA max
(1)
450
150
450
50
-
350
-
450
-
-
350
450
50
-
450
-
450
-
-
PDI min
-450
-50
-
-
-100
-
-
-100
-
-
-100
-
-
PDI max
450
450
-
-
450
-
-
450
-
-
450
-
-
Def =
Default value of a parameter settable by user.
Min =
Minimum settable value or possible value.
Max =
Maximum settable value or possible value.
Auto =
Limit is calculated by software and is not settable by user.
BP
= Blood pump.
(1) PDPA max: 450 mmHg is the limit in the first 20 sec after fluid side pumps start running.
(2) PBE min, max: for more details see above
(3)
PV min, max: for more details see above