Operating Manual
ALARMS AND PROBLEM CORRECTION
Software 2.6
B|BRAUN
Chapter 8
Page 5 of 27
SHARING EXPERTISE
2.6_CA-EN_03
Code
Alarm Text
Alarm Cause
Corrective Action
A 26
Weight test 1 error - Is
PCLD chamber full ?
It was determined with
weight test 1 that the
plasma/buffer pump does
not deliver correctly.
Malfunction of the
plasma/buffer pump
Malfunction of the
load cell
Ensure that:
The seal on the saline bag is open.
The clamp on the buffer line is open.
The buffer line is not kinked or clamped.
The plasma/buffer pump segment is not
inserted crosswise and in the correct
direction.
After eliminating the the cause of the alarm
and acknowledging the alarm, the test is
automatically repeated.
A 27
Weight test 2 error
It was determined with
the weight test 2 that the
dialysate pump does not
deliver correctly.
Dialysate flow
obstructed
Malfunction of the
load cell
Ensure that:
The seals of the dialysate bags are open.
The clamps of the dialysate lines are open.
The dialysate line is not kinked or clamped.
The bags are hanging motionless on the
load cell.
After eliminating the the cause of the alarm
and acknowledging the alarm, the test is
automatically repeated.
A 28
DAD test failed
An error occurred during
the DAD check.
Malfunction of DAD
Ensure that:
The dialysate line is inserted in the air
detector for the dialysate (DAD).
The clamps on the dialysate line are open.
The seals of the dialysate bags are open.
The connections between the dialysate bags
and the dialysate line are firmly seated.
After eliminating the the cause of the alarm
and acknowledging the alarm, the test is
automatically repeated.
A 29
Pressure test failed
Pressure build-up and
pressure holding test
failed
Ensure that:
The PBE pressure transducer is screwed on
correctly.
The venous line was inserted in the safety
air clamp (SAK).
All lines were installed according to
instruction.
The venous pressure transducer (PV) is
correctly screwed on.
A 30
Leak test failure /
venous lead inserted in
SAK ?
An error occurred during
the check of the safety air
clamp (SAK) and the line
leakage test.
Ensure that:
The venous line is inserted in the safety air
clamp (SAK).
The connections between the lines and the
filters are firmly seated.
The venous pressure transducer (PV) is
correctly screwed on.
After eliminating the the cause of the alarm
and acknowledging the alarm, the test is
automatically repeated
.