Operating Manual
ALARMS AND PROBLEM CORRECTION
Software 2.6
B|BRAUN
Chapter 8
Page 27 of 27
SHARING EXPERTISE
2.6_CA-EN_03
8.3.7 Changing the H.E.L.P. Ultrafilter
Accessories
Article number
H.E.L.P. Ultrafilter SMC 1,8
7210932
1 x collection bags
7210929
3 l H.E.L.P. 0.9% NaCl solution
4107
2 connection lines
7210934
Attach a connection line with the NaCl solution, fill
the line and connect it with the red, plasma-side
filter opening.
Attach the second connection line and the empty
bag as shown in the Figure with the blue, plasma-
side filter opening.
Hold the filter so that it is filled from the bottom to
the top and thoroughly vented in the process.
Clamp shut both connection lines when approx. 1 l
rinse solution has flown through (be careful that no
air enters the filter!).
Switch the machine to bypass mode by selecting
<Stop Priming
> or <
Stop Therapy
> in the menu
bar and confirm with the
key.
Clamp shut the connection and reinfusion line
leading to the dialyzer, remove the old filter and
connect the new filter in the correct orientation to
the connection and reinfusion lines. Connect the old
filter with the connection lines on rinse solution and
drain bag.
Plug the Hansen connectors from the old onto the
new filter (hold old filter horizontally!). Observe the
color marking. Insert the new filter with the blue
end down into the support.
Fill the dialysate side of the filter by manually
rotating the dialysate pump.
Reopen the connection and reinfusion lines and
continue the interrupted phase by selecting
<Start
Priming
> or <
Start Therapy
> and confirm with the
key.
3 l
0.9% NaCl
Empty
bag