Operating Manual
THERAPY
Software 2.6
B|BRAUN
Chapter 5
Page 8 of 14
SHARING EXPERTISE
2.6_CA-EN_03
Press the
key to select the parameter to be
changed. The field is shown with a red background
and white labeling. Perform the desired change using
the rotary knob and confirm with the
key.
The changing of the following parameters must be
confirmed with the
key since they are relevant
to patient safety:
Plasma flow rate %
Plasma volume
Balance
Heparin bolus
Heparin flow rate
If a parameter is relevant to patient safety, the
currently set value is shown in the Setting window
above the setting range. In addition, the LEDs above
the
key blink.
To quit the screen for setting the parameters, press
the
key. The cursor changes back to the menu
bar and the menu item <
Start Therapy
> of the
Parameter Overview screen.
If no settings are changed for more than 15 seconds,
the screen automatically changes back to the
previously set screen.
For more details, see chapter 4.2.1
5.3.2 Parameter Setting in the <Parameter
Overview> Screen
Using the rotary knob and the
key, change to
the <
Parameter Overview
> screen.
OK
OK
OK
W16: Press 'OK' to return to menu selection
Setting
Plasma flow
%
20
[ 10 : 40 ]