Operating Manual
PRIMING AND RINSING
Software-Version 2.6
B|BRAUN
Chapter 4
Page 16 of 19
SHARING EXPERTISE
2.6_CA-EN_03
In addition to the parameters listed in chapter 4.2.1.,
the following parameters can be entered:
Temperature
Default setting:
39
o
C
Range:
34 ÷ 40
o
C
Step size:
0.5
o
C
PA min
Default setting:
-150 mmHg
Range:
-250 ÷ 80 mmHg
Step size:
10 mmHg
PA max
Default setting:
100 mmHg
Range:
0 ÷ 200 mmHg
Step size:
10 mmHg
PV window Min
Default setting:
20 mmHg
Range:
10 ÷ 40 mmHg
Step size:
5 mmHg
PV window Max
Default setting:
40 mmHg
Range:
20 ÷ 100 mmHg
Step size:
5 mmHg
PPL min
Default setting:
-10 mmHg
Range:
-20 ÷ 10 mmHg
Step size:
1 mmHg
TMP max
Default setting:
100 mmHg
Range:
20 ÷ 200 mmHg
Step size:
10 mmHg
PPF min
Default setting:
-20 mmHg
Range:
-50 ÷ 50 mmHg
Step size:
5 mmHg
PDF min
Default setting:
-50 mmHg
Range:
-50 ÷ 0 mmHg
Step size:
5 mmHg
PDF max
Default setting:
350 mmHg
Range:
10 ÷ 400 mmHg
Step size:
10 mmHg
PDPA max
Default setting:
150 mmHg
Range:
50 ÷ 350 mmHg
Step size:
10 mmHg
PPL Threshold
Default setting:
20 mmHg
Range:
-10 ÷ 120 mmHg
Step size:
5 mmHg