Operating Manual
PRIMING AND RINSING
Software-Version 2.6
B|BRAUN
Chapter 4
Page 10 of 19
SHARING EXPERTISE
2.6_CA-EN_03
Press the
key to select the parameter to be
changed. The field is shown with a red background
and white labeling. Perform the desired change using
the rotary knob and confirm it with the
key.
The changing of the following parameters must be
confirmed with the
key since they are relevant
to patient safety:
Plasma flow rate
Plasma volume
Balance
Heparin bolus
Heparin flow rate
If a parameter is relevant to patient safety. the
currently set value is shown in the Setting window
above the setting range. In addition, the LEDs above
the
key blink.
To quit the screen for setting the parameters, press
the
key. The cursor changes back to the menu
bar of the Main Parameter screen and the menu item
<
Start Priming
>.
If you do not perform any settings for more than 15
seconds, the screen automatically changes back to
the previously selected screen.
The following parameters can be set in the priming
and rinsing phase:
Plasma flow
Default setting:
20 % of blood flow
Range:
10 ÷ 40 % of blood flow
Step size:
1 % of blood flow
The plasma flow is limited to a maximum of 40 % of
the blood flow, and 50 ml/min.
If the blood flow is changed manually, the plasma
flow is automatically changed according to the set
ratio.
The plasma flow is set in % of blood flow, and is
displayed in ml/min.
OK
OK
OK
Setting
Plasma flow
%
20
[ 10 : 40 ]
W16:
Press 'OK' to return to menu selection!