Operating Manual
THERAPY
Software 2.6
B|BRAUN
Chapter 5
Page 10 of 14
SHARING EXPERTISE
2.6_CA-EN_03
Press the
key to select the parameter to be
changed. The field is shown with a red background
and white labeling. Perform the desired change using
the rotary knob and confirm it with the
key.
The changing of the following parameters must be
confirmed with the
key since they are relevant
to patient safety:
Plasma flow rate
Plasma volume
Balance
Heparin flow rate
Heparin bolus
PA min
PA max
PV MIN window (mmHg)
PV MAX window (mmHg)
Ratio Dialysate/Plasma
If a parameter is relevant to patient safety, the
currently set value is shown in the Setting window
above the setting range. In addition, the LEDs above
the
key blink.
WARNING
Risk to patient due to blood loss since increasing
the PV MIN window elevates the likelihood of an
unrecognised removal of the venous access.
Do not cover the venous access.
Keep the patient under continuous
surveillance.
To quit the screen for setting the parameters, press
the
key. The cursor changes back to the menu
bar and the menu item <
Start Therapy
> of the Main
Parameter screen.
If no settings are changed for more than 15 seconds,
the screen automatically changes back to the
previously set screen.
For more details, see chapter 4.2.2.
OK
OK
OK
W16: Press 'OK' to return to menu selection
Setting
Plasma flow
%
[ 10 : 40 ]