Operating Manual
Installation and Technical Data
Software 2.6
B|BRAUN
Chapter 9
Page 6 of 17
SHARING EXPERTISE
2.6_CA-EN_03
9.2.3 Service and Maintenance
Repairs and maintenance may be performed only by
personnel authorized and trained by the manufacturer.
No special maintenance by the user is required.
The Technical Safety Inspection is to be performed
every twelve months based on the Service Manual and
the Operating Manual, subject to technical changes,
and to be documented.
The maintenance of the calibration sensors (load cell,
temperature, pressures, blood leak detector, SAD etc.)
must be performed in accordance with the
specifications of the Service Manual and the
respective working instructions.
If the exchanging of fuses is required, only the fuses
specified by the manufacturer may be used (see
Service Manual).
9.2.4 Disposables, Consumables and Accessories /
Replacement Parts
The machine may be used only in combination with
the H.E.L.P. apheresis treatment system.
When using the approved single-usage articles,
consumables and accessories, observe the instructions
for use of the respective components.
Dispose of the single-usage articles required for the
treatment according to the local regulations.
Use only accessories and replacement parts whose
suitability with respect to technical safety has been
established and certified by an inspection authority.
This verification must be carried out by an inspection
authority who is authorized to inspect the ready-for-
use machine.
9.2.5 Cleaning and Disinfection
All modules of the H.E.L.P. Plasmat® Futura and the
screen may be cleaned with ethanol-containing (
70%) or isopropano-containing (
60%) surface
disinfectants. Please observe the instruction for use of
the respective manufacturer