Operating Manual
Installation and Technical Data
Software 2.6
B|BRAUN
Chapter 9
Page 16 of 17
SHARING EXPERTISE
2.6_CA-EN_03
Patient balance
Patient balance feedback control system based on weight
measurement by the load cell controlling the reinfusion pump
(marked green).
Patient balance range:
- 600 ÷ 0
g
Allowed tolerance
± 50
g
Working range:
- 600 ÷ 0
g
During Therapy:
Alarm limits:
± 200
g
Patient balance (calculated by the software from weight
change) is compared continuously to the momentary
theoretical value.
Protection system:
Double channel patient balance monitoring with sensor
test during preparation phase.
Alarm override:
Alarm limit can be increased by 100 g by alarm
acknowledge, but reaching the alarm limit ± 400 g
override is not possible anymore.
Acoustic alarm interval:
120
s
9.3.8 Estimation of Required Number of Dialysate Bags
Based on both the ratio dialysate/plasma and the required plasma volume, the following table shows
an estimation of the required number of dialysate bags.
Ratio Dialysate/Plasma
Plasma Volume
4
5
6
7
8
9
10
11
12
3000
3
4
5
6
6
7
8
9
9
3500
4
5
6
6
7
8
9
10
11
4000
4
5
6
7
8
9
10
11
12
4500
5
6
7
8
9
10
11
12
13
5000
5
7
8
9
10
11
12
14
15
5500
6
7
8
10
11
12
13
15
16
6000
6
8
9
10
12
13
15
16
17