ALARMS AND PROBLEM CORRECTION
Software 2.6
B|BRAUN
Chapter 8
Page 12 of 27
SHARING EXPERTISE
2.6_CA-EN_03
Code
Alarm Text
Alarm Cause
Corrective Action
A 95
(S) Line in SAD not
empty or SAD error
Consumables already or
still mounted
Remove all consumables from the machine.
A 96
(S) Load cell not empty
or load cell error
Consumables already or
still mounted
Remove all consumables from the machine.
A 97
(S) Venous pressure (PV)
not zero !
Consumables already or
still mounted
Remove all consumables from the machine.
A 98
(S) Arterial pressure (PA)
not zero !
Consumables already or
still mounted
Remove all consumables from the machine.
A 99
(S) Control system not
working properly !
Erroneous controller or
user interface function
Acknowledge the alarm (twice). If it is not
possible switch the machine off and on again
to eliminate a possible transient failure.
If the problem cannot be solved, close the
treatment immediately and inform technical
service.
A 100
(S) SAD clock test failed.
Switch off and on
Erroneous SAD clock
function
Switch the machine off and on.
If alarm remains after power off, call service.
A 103
(S) Balance error
Balance error > 500 g
Defect of plasma/buffer
pump, of reinfusion pump
or of load cell
Ensure that:
The seals of the saline bags and of the
dialysate bags are open.
The lines are not kinked or clamped.
The clamps on the buffer line and on the
dialysate line are open.
The dialysate line is inserted into the
support on the load cell.
The pump segments are correctly inserted.
A 104
(S) Plasma volume error Count error of Plasma
treated volume
Ensure that:
The plasma lines are not kinked or clamped.
The pump segments are correctly inserted.