Operating Manual
PRODUCT DESCRIPTION
Software 2.6
B|BRAUN
Chapter 2
Page 9 of 15
SHARING EXPERTISE
2.6_CA-EN_03
2.4 MONITOR
2.4.1 Monitor Controls
The
rotary knob
moves the cursor on the screen.
Display in lines:
Clockwise rotation - the cursor moves from left to
right
Counterclockwise rotation - the cursor moves from
right to left
Display in columns:
Clockwise rotation - the cursor moves from top to
bottom
Counterclockwise rotation - the cursor moves from
bottom to top
The set parameters are accepted by pressing the
key
.
The
key
confirms important actions, such as
Phase change (e.g. change from the
priming/rinsing phase to the therapy phase).
Quitting the <
Parameter Setting
> menu.
Acknowledging messages that require immediate
action (e.g. prompt for turning over the dialyzer
during the priming and rinsing phase).
When this key is active, the
yellow LEDs
above it
light. These LEDs blink during adjustment of
parameters with relevance to patient safety.
When an acoustic alarm occurs, switch off the alarm
with the
key
. After elimination of the cause of
the alarm, acknowledge the alarm with the
key
and continue with the respective phase.
When this key is active, the
red LEDs
above it light.
Alarms initiated by opening a pump cover are
reset by closing the pump cover.
Keys for operating the blood pump
The
key
and the
key
lower or raise the
delivery rate of the blood pump.
If the blood pump stops during an alarm, the
red
LEDs
light. If the blood pump runs, the
green LEDs
light. If both LEDs blink alternately, the blood pump
OK
Rotary knob
Enter key
OK key with LEDs
Alarm key with LEDs
Key for operation of
the blood pump
(-) key
(+) key
Red LEDs
Green LEDs
Start/Stop key