Operating Manual
Installation and Technical Data
Software 2.6
B|BRAUN
Chapter 9
Page 15 of 17
SHARING EXPERTISE
2.6_CA-EN_03
Plate warmer (H
)
Fluid warming system with temperature sensors based on heat
transfer between temperature controlled metal plate and plastic
dialysate bag
Temperature range:
34.0 ÷ 40.0
°C
Default in therapy:
39.0
°C
Allowed variation:
0.5
°C
Upper alarm limit:
41.5°C for 10 seconds.
Protection system:
Double channel temperature monitoring with sensor test
during preparation phase.
Alarm override:
Not possible during the therapy
Acoustic alarm interval:
120
s
9.3.7 Weight System
Load cell
Loading capacity:
30
kg
Weight resolution:
1
g
Linearity:
0.015
%
Working range:
0.00 ÷ 25.00
kg
Overload protection:
Electrically at 24.5
kg
Mechanically at 26.0
kg
Weight change alarm:
Weight deviation < 50 g:
no alarm
Weight deviation 50÷200 g: alarm after 5 seconds if
deviation has not been
corrected
Weight deviation > 200 g:
immediate alarm
Protection system:
Sensor test during preparation phase and electric current
through load cell bridge monitoring during therapy.
Alarm override:
Not possible during the therapy.
Acoustic alarm interval:
120
s