Operating Manual
ANNEX 3 - LIMIT VALUES
Software 2.6
B|BRAUN
Annex 3
Page 2 of 2
SHARING EXPERTISE
2.6_CA-EN_03
Limits of Adjustable Parameters
Parameter
Default
min
max
Step
Sequence
Unit
Blood flow
40
10
150
5
ml/min
Plasma flow
20
10
40
1
% blood flow
Plasma reinfusion volume
400
400
1000
50
ml
Blood reinfusion volume
300
100
600
50
ml
Return flow
30
10
50
5
ml/min
Ratio Dialysate/Plasma
4
4
12
1
Rinsing volume
2400
2400
20000
100
ml
Plasma volume
3000
100
6000
50
ml
Patient Balance
0
-600
0
50
g
Temperature
39
34
40
0,5
°C
Heparin flow
2,0
0,0
10,0
0.5
ml/h
Heparin bolus
1,0
0
10
0,5
ml
Autostop heparin
0
0
60
5
min
PA min
-150
-350
80
10
mmHg
PA max
100
0
200
10
mmHg
Minimum PV window
20
10
40
5
mmHg
Maximum PV window
40
20
100
5
mmHg
PPL min
-10
-20
10
1
mmHg
PPL threshold
20
-20
(1)
120
5
mmHg
TMP max
100
20
200
10
mmHg
PPF min
-20
-50
50
5
mmHg
PDF min
-50
-50
0
5
mmHg
PDF max
350
10
450
10
mmHg
PDPA max
150
50
350
10
mmHg
(1) Default PPL threshold (min): -10 mmHg