Operating Manual
REINFUSION
Software 2.6
B|BRAUN
Chapter 6
Page 10 of 14
SHARING EXPERTISE
2.6_CA-EN_03
To set the parameters, select the <
Parameter
Setting
> menu item with the cursor in the
<Parameter Overview
> screen and activate it with
the
key.
All parameters which can be changed are displayed in
red. The currently selected parameter has a gray
background. The Setting window displays the
allowable range. Using the rotary knob, you can select
the individual parameters.
The following parameters can be set in the reinfusion
phase:
Reinfusion flow (ml/min)
Temperature (
o
C)
PA min (mmHg)
PA max (mmHg)
PV MIN window (mmHg)
PV MAX window (mmHg)
PPL min (mmHg)
TMP max (mmHg)
PPF min (mmHg)
PDF min (mmHg)
PDF max (mmHg)
PDPA max (mmHg)
PPL Threshold (mmHg)
Ratio Dialysate/Plasma
W16: Press 'OK' to return to menu selection!
Setting
Return flow
ml/min
30
[ 10 : 50 ]