Operating Manual
PREPARATION
Software 2.6
B|BRAUN
Chapter 3
Page 7 of 9
SHARING EXPERTISE
2.6_CA-EN_03
(1)
Mount the plasma filter in its holder.
(2)
Connect the plasma line including the reinfusion
line to the top right of the plasmafilter.
(3)
Connect the venous line to the top left of the
plasma filter.
(4)
Connect the reinfusion line to the 1500 ml
compartment of the double chamber bag with
the saline solution. Break the seal and fill the
line manually until the saline solution reaches
the plasma line. Then close the clamp at the
reinfusion line.
(5)
Mount the heparin adsorber in its holder and
connect the inlet line and outlet line.
(6)
Place the feeder line to the heparin adsorber into
the automatic clamp HAK. Make sure that the
line is correctly inserted in the clamp.
(1)
Place the venous line into the safety air detector
SAD.
(2)
and into the safety air clamp SAK.
(3)
Connect the venous line to the 5L-empty bag
which is hanging on the IV-pole.
6
1
2
3
1
2
3
4
5