Operating Manual
ALARMS AND PROBLEM CORRECTION
Software 2.6
B|BRAUN
Chapter 8
Page 17 of 27
SHARING EXPERTISE
2.6_CA-EN_03
Code
Warning Text
Reason for Warning
Corrective Action
W 33
Heparin bolus … ml.
Safety query before
administering the set
heparin bolus
Press the
key to administer the
heparin bolus.
If you do not wish to administer the
heparin bolus, wait 5 s for the Warning
window to disappear.
M
W 35
Activate reinfusion ?
Prompt for changing
to reinfusion mode
Press the
key to change to the
reinfusion phase.
M
W 36
Are you sure to start a new
therapy?
Return to this therapy is
not possible.
Information before
returning to the Start
screen.
Press the
key if you wish to return
to the Start screen.
M
Note that the data of the currently performed therapy
are deleted when you return to the Start screen.
W 37
Selftests completed, check
characters and press ENTER
Confirmation of the
successfully performed
initial selftest
Select ‘END’ softkey and press
.
T
W 39
Power fail eliminated !
Check lines, filters and
parameter setting, restart !
Information after a
power failure
Press the
key after verification of
the required positions to continue
therapy.
M
W 41
Open plasma clamp and
close venous clamp!
Confirmation in the
middle of blood
reinfusion (after
150 ml)
Press the
key after opening/closing
the respective clamps to continue blood
reinfusion.
M
W 42
Set Plasma Flow is too low.
Increase Blood or Plasma
Flow.
Information that the
required Plasma Flow
is too low (< 2 ml/min)
Increase the Blood Flow or increase the
Plasma Flow value.
T
W 43
Attention! Precipitate filter
rupture possible!
Check PPF chamber level,
PPF protector and
connection
Or check air bubbles in
chamber and chamber
locking.
PPF chamber level is
too high, PPF protector
is wet. No PPF
pressure increase in
case of closed HAK
clamp.
(This warning appears
together with alarm
A73)
Ensure that:
The PPF chamber level is not too high
and PPF protector is not wet.
The PPF protector is connected properly.
The PPF chamber is positioned and the
level sensor is locked properly.
No air bubble is attached to the inner
chamber wall.
Then press the
key after
examination to continue therapy.
M
W 44
W44: Patient Balance too
high or Plasma Flow too
low. Please adjust.
The required Patient
Balance cannot be
reached in the
remaining therapy
time. Balance error
might occur later
during the course of
the treatment.
Reduce the Patient Balance value or
increase the Plasma Volume value or
increase Plasma Flow value.
T
OK
OK
OK
OK
OK
OK