Operating Manual
PREPARATION
Software 2.6
B|BRAUN
Chapter 3
Page 6 of 9
SHARING EXPERTISE
2.6_CA-EN_03
3.4 SETTING UP THE H.E.L.P. FUTURA SET
(1)
Place the plastic plate of the H.E.L.P. Futura kit
on to the lower support on the machine.
Press the plate against the front of the machine.
(2)
Secure the plate with the upper rotary
attachment knob (
2
).
(1)
Place the pump segment of the reinfusion line
into the reinfusion pump (marked green).
(2)
Place the pump segments of the plasma/buffer
line successively into the plasma/buffer pump
(marked brown and yellow).
(3)
Place the plasma line coming from the plasma
filter into the blood leak detector BLD.
(4)
Check whether the pump segments are inserted
in the correct orientation.
(1)
Place the two air chambers into the holders as
shown. Lock them in place in the holder by
turning the black lock.
(2)
Screw on the four pressure transducers as
shown.
(3)
Place the venous air chamber into the holder
provided and screw on the venous pressure
transducer as shown.