Operating Manual
THERAPY
Software 2.6
B|BRAUN
Chapter 5
Page 3 of 14
SHARING EXPERTISE
2.6_CA-EN_03
Balance in g
Arterial pressure in mmHg
Venous pressure in mmHg
Prefilter pressure in mmHg
Plasma pressure in mmHg
Select <
Start Therapy
> in the menu bar. The
following message is displayed in the Warning
window: <
W15: Connect buffer – seal and
clamp opened?>
Exchange the saline bag on the load cell with the
prepared acetate buffer bag.
Remove the venous line from the empty bag on
the IV-pole and screw it to the second connection
of the saline bag on the IV-pole (next to the
arterial line).
Remove the empty bag from the IV-pole.
Remove the clamps from the bag and the buffer
line and make sure that all bag break seals are
open.
At this point at the latest, enter the parameters
required for the therapy, such as plasma volume,
heparin flow, heparin bolus, etc. (see chapter 4.2).
Confirm the message in the Warning window with the
key.
The machine is now ready for the therapy and can be
connected to the patient.
Starting the Blood Circuit
Disconnect the arterial line from the physiological
saline bag on the IV-pole.
Connect the line to the patient access for drawing
blood.
The green and red LEDs above the
key blink
alternately. Start the blood pump with the
Start
Stop
Start
Stop
OK
Warning
W15: Connect buffer – seal and clamp opened ?
Press ‘OK’ to proceed !