Operating Manual
PRIMING AND RINSING
Software-Version 2.6
B|BRAUN
Chapter 4
Page 5 of 19
SHARING EXPERTISE
2.6_CA-EN_03
The prompt <
W18: Break seals and open all
clamps!
> appears in the Warning window.
Press the
key to continue.
The <
Start Priming
> command in the menu bar
blinks (the label changes between black and
gray). This shows that an input is expected from
the user.
After starting priming by selecting <
Start Priming
>,
the message <
W01: Plasma pump starts after
pressurizing blood side
> is displayed in the message
line.
Automatic Filling of Blood Side
During automatic filling, the arterial line, the plasma
filter and the venous line are rinsed and filled by
default with 600 ml saline solution.
Start filling the arterial line by pressing the
key.
Step 1/2
The arterial line, the plasma filter and the venous line
are filling. The preset blood flow rate is 150 ml/min.
Step 3
The safety air clamp (SAK) opens and then closes
again and the level of the arterial chamber is set
accordingly. This vents the plasma filter.
OK
W01: Plasma pump starts after pressurizing blood side
Warning
W18: Break seals and open all clamps !
Press ‘OK’ to proceed !