Operating Manual
ALARMS AND PROBLEM CORRECTION
Software 2.6
B|BRAUN
Chapter 8
Page 22 of 27
SHARING EXPERTISE
2.6_CA-EN_03
8.3.3 Changing the Solution Bags
Change as a result of a defective bag
Select the
<Stop Therapy
> function to go to the
bypass mode (blood pump is turning, plasma-side
pumps stand still).
Attach a clamp to the bag to be exchanged and
close the clamp on the feed line.
Exchange the defective bag for a new bag.
Break the seal of the new bag.
Open the clamp of the feed line again.
Confirm the warning W09 <Check lines and bags>
by pressing the
key.
Continue the treatment by selecting the <
Start
Therapy
> function.
Change at a treatment volume > 3000 ml
At a treatment volume > 3010 ml, the H.E.L.P.
Plasmat® Futura automatically switches to bypass. The
warning <
W 10: Plasma vol. > 3 l. Change buffer
bag and check dialysate bags
> is displayed. Remove
the full drain bags and replace them.
Attach a clamp to the feeding buffer line.
Remove the empty acetate buffer bag and replace
it by a new prepared acetate buffer bag.
Open the seal of the new acetate buffer bag.
Reopen the clamp on the buffer line again.
Check also whether sufficient dialysate is available
and replace dialysate bags if necessary.
Confirm the change by pressing the
key.
Continue the therapy by selecting the
<Start
Therapy
> function.
Change of the dialysate bags if they are nearly
empty
If the ratio dialysate/plasma is > 1:4 and the dialysate
bags are nearly empty, the H.E.L.P. Plasmat® Futura
automatically switches to bypass. The warning
<
W 45: Dialysate bags nearly empty. Change bags
if necessary.
> is displayed.
a) Exchange dialysate bags if more dialysate solution is
required:
Attach a clamp to the feeding dialysate line.
Remove the empty dialysate bag and replace it
by a newly prepared dialysate bag.
Open the seal of the new dialysate bag.
Reopen the clamp on the dialysate line again.
Repeat for the other dialysate bags if
OK