Operating Manual
SAFE HANDLING
Software-Version 2.6
B|BRAUN
Chapter 1
Page 8 of 10
SHARING EXPERTISE
2.6_CA-EN_03
damaged in any way. A damaged H.E.L.P. Plasmat®
Futura machine must be submitted for repairs or
disposed of.
1.6 INTERACTION WITH OTHER DEVICES
It is recommend that the machine be connected to a
dedicated circuit. When using the H.E.L.P. Plasmat®
Futura machine in combination with other
therapeutic devices, it is recommended to use a
c
onnection line for electrical ground
, since the
leakage currents from all connected devices are
additive.
1.6.1 Electromagnetic Interaction
The H.E.L.P. Plasmat® Futura machine was developed
and tested in accordance with the valid standards for
interference suppression and EMC. It cannot, however,
be guaranteed that no electromagnetic interaction
with other devices will occur.
Examples: mobile phones, computer tomograph (CT)
It is recommended that the use of mobile phones and
other devices emitting strong electromagnetic
radiation be restriced to a minimum distance from the
H.E.L.P. Plasmat® Futura machine (refer to table in
Chapter 9).
Operation of other therapeutic or diagnostic medical
devices in conjunction with the H.E.L.P. Plasmat®
Futura machine, or use of non medical devices
directly near the H.E.L.P. Plasmat® Futura machine,
should be carefully monitored.
1.7 INFORMATION FOR THE OPERATOR
1.7.1 Training by Manufacturer Prior to
Commissioning
The operator may only use the device after the
manufacturer has trained the responsible staff based
on these instructions for use.