Operating Manual
THERAPY
Software 2.6
B|BRAUN
Chapter 5
Page 2 of 14
SHARING EXPERTISE
2.6_CA-EN_03
5.1 Starting the Therapy
After the completion of the priming and rinsing
phase, select the <
Therapy
> menu item in the
menu bar and confirm with the
key
The following message is displayed in the warning
window
<W32: Activate therapy mode ?>
Confirm the message with the
key.
The change to the therapy phase is possible only when
the minimum rinsing volume of 2400 ml has been
reached.
The screen changes to the Therapy screen.
Display Area of Therapy Screen
Blood flow in ml/min
Heparin flow in ml/h
Heparin bolus in ml
Autostop heparin in min
Plasma flow in ml/min
Therapy time [Actual/Rest] in hh:mm
Plasma volume [Actual/Rest] in ml
OK
Warning
W32: Activate therapy mode ?
Press ‘OK’ to proceed !
Warning