Operating Manual
ALARMS AND PROBLEM CORRECTION
Software 2.6
B|BRAUN
Chapter 8
Page 19 of 27
SHARING EXPERTISE
2.6_CA-EN_03
8. 3 PROBLEM CORRECTION
8.3.1 Balance Reset
Balance error > 200 g
For a balance error > 200 g, the alarm <
A69: Balance
error!
> and the warning <
W28: Balance error
> are
displayed. Check whether:
The bags are hanging correctly on the load cell.
All seals and clamps are open.
All lines are free from kinks.
Acknowledge the alarm with the
and the
keys after you have eliminated the cause of the error.
The warning <
W28: Balance error
> is displayed until
the balance error has been compensated.
Balance error > 300 g
If the balance error remains and exceeds a value of
300 g, the alarm <
A69: Balance error !
> is initiated
and the warning <
W24: Balance error > 300 g,
check lines and bags !
> displayed.
Check the system as described above. Acknowledge the
alarm and the warning with the
and the
keys after you have eliminated the cause of the error.
The warning <
W28: Balance error
> is displayed until
the balance error has been compensated.
Balance error > 400 g
If it was not possible to correct the balance error with
the measures described above and it exceeds a value
of 400 g, the alarm <
A69: Balance error !
> is
initiated again and the warning <
W25: Balance error
> 400 g END OF THERAPY IS RECOMMENDED
> is
displayed.
The end of therapy is recommended to exclude a
balance error in the fluid balance of the patient.
Terminate the therapy as described in chapter 6.
OK
OK