Operating Manual
ALARMS AND PROBLEM CORRECTION
Software 2.6
B|BRAUN
Chapter 8
Page 14 of 27
SHARING EXPERTISE
2.6_CA-EN_03
8.2.2 List of Warnings
M = display in the Message box, T = display in the
Alarm/Note line
Code
Warning Text
Reason for Warning
Corrective Action
W 01
Plasma pump starts after
pressurization blood side
Indication that the
arterial line is filled
and the filling phase is
continuing.
T
W 03
Press 'OK' to confirm safety
data
Safety query when
parameters with safety
relevance have been
changed
The changed parameters have safety
relevance. Check the setting thoroughly
and confirm with the
key.
M
W 04
Turn dialyzer (blue side
down) !
In the filling phase,
the next handling step
is indicated.
Turn over the dialyzer and confirm with
the
key
M
W 05
Therapy stop for too long !
Therapy interrupted
for more than 5
minutes
Continue the therapy.
Select the
<Start Therapy
> command
and confirm with the
key.
T
W 06
Therapy completed !
The end of the therapy
is indicated.
Press the
key to change to the
reinfusion phase.
M
W 08
Reinfusion stop for too
long !
Reinfusion interrupted
for more than 5
minutes
Continue the reinfusion.
Select the
<Start Reinfusion
> command
and confirm with the
key.
T
W 09
Check lines and bags !
Deviation of total
weight on the load cell
in bypass
Check the bags and lines and perform the
necessary corrections.
Press the
key to continue.
M
W 10
Plasma vol. > 3 l.
Change buffer bag and
check dialysate bags
Solution volume not
sufficient for
continuation of the
treatment.
Remove the buffer bag and hang on a
new one.
If necessary, remove the full drain bags
and hang on new dialysate bags.
Press the
key to continue.
For changing the solution bags, see
chapter 8.3.3.
M
W 11
1) Connect reinfusion and
buffer lines to saline
solution
2) Clamp plasma line at out
of plasma filter
3) Turn plasma and
precipitate filters
4) Turn heparin adsorber
Information for
preparing the plasma
reinfusion
Follow the instructions on the monitor
and then press the
key to continue.
M
OK
OK
OK
OK
OK
OK