Precipitate filter pressure (PPF)
Electronically measured by a pressure sensor and digitally
displayed
Measurement range:
- 500 ÷ +500
mmHg
Allowed tolerance
± 10
mmHg
Working range:
+150 ÷ +300
mmHg
During Therapy:
Default alarm limits:
-20 ÷ +450
mmHg
Lower limit adjustable in parameter setting
Protection system:
Sensor test during preparation phase.
Alarm override:
Not possible during the therapy
Acoustic alarm interval:
120
s
Dialyzer pressure (PDF)
Electronically measured by a pressure sensor and digitally
displayed
Measurement range:
- 500 ÷ +500
mmHg
Allowed tolerance
± 10
mmHg
Working range:
+120 ÷ +270
mmHg
During Therapy:
Default alarm limits:
-50 ÷ +350
mmHg
Adjustable in parameter setting
Protection system:
Sensor test during preparation phase.
Alarm override:
Not possible during the therapy
Acoustic alarm interval:
120
s
Reinfusion pump (PRP)
marked green
Peristaltic roller pump with motor switch-off when the pump
cover is open
Delivery rate:
1 ÷ 60
ml/min
Controlled by patient balance feedback control system
(based on weight measurement by load cell).
Allowed tolerance:
< 10
%
Working pressure range:
-100 ÷ +450
mmHg
Protection system:
Pump status and rate is monitored via a rotation detector.
Alarm override:
Not possible during the therapy
Acoustic alarm interval:
120
s