Operating Manual
Installation and Technical Data
Software 2.6
B|BRAUN
Chapter 9
Page 9 of 17
SHARING EXPERTISE
2.6_CA-EN_03
9.3.4 Extracorporeal Blood Circuit
Blood pump (BP)
Peristaltic roller pump with motor switch-off when the pump
cover is open
Delivery rate:
10 ÷ 150
ml/min
Delivery rate tolerance:
< ± 10
%
Working pressure range:
-140 ÷ +500
mmHg
Protection system:
Pump status and rate is monitored via a rotation detector.
Alarm override:
Not possible during the therapy
Acoustic alarm interval:
120
s
Arterial pressure (PA)
Electronically measured by a pressure sensor and digitally
displayed
Measurement range:
- 500 ÷ +500
mmHg
Allowed tolerance
± 10
mmHg
Working range:
-60 ÷ +10
mmHg
During Therapy:
Default alarm limits:
-150 ÷ +100
mmHg
Adjustable in parameter setting
Protection system:
Double channel pressure monitoring with sensor test
during preparation phase.
Alarm override:
Not possible during the therapy
Acoustic alarm interval:
120
s
Prefilter pressure (PBE)
Electronically measured by a pressure sensor and digitally
displayed
Measurement range:
- 500 ÷ +500
mmHg
Allowed tolerance
± 10
mmHg
Working range:
+90 ÷ +140
mmHg
During Therapy:
Alarm limits:
-140 ÷ +250
mmHg
Default alarm window:
Automatic control
Lower limit:
Reference value - 40
mmHg
Upper limit:
Reference value + 80
mmHg
Protection system:
Sensor test during preparation phase.
Alarm override:
Not possible during the therapy
Acoustic alarm interval:
120
s