Operating Manual
REINFUSION
Software 2.6
B|BRAUN
Chapter 6
Page 11 of 14
SHARING EXPERTISE
2.6_CA-EN_03
Press the
key to select the parameter to be
changed. The field is shown with a red background and
white labeling. Perform the desired change using the
rotary knob and confirm it with the
key.
The changing of the following parameter must be
confirmed with the
key since it is relevant to
patient safety:
Reinfusion flow in ml/min
PA min in mmHg
PA max in mmHg
PV MIN window (mmHg)
PV MAX window (mmHg)
Ratio Dialysate/Plasma
Plasma and blood reinfusion volume can be set only in
the <
Main Parameter
> screen.
Parameters relevant to patient safety can be seen in
the Setting window. The currently set value is shown
above the setting area. In addition, the LEDs above the
key blink.
To quit the screen for setting the parameters, press the
key. The cursor changes back to the menu item
<
Start Plasma
> of the menu bar of the Parameter
Overview screen.
If no settings are changed for more than 15 seconds,
the screen automatically changes back to the
previously set screen.
6.4.3 Parameter Setting in the <Flow Scheme>
Screen
Using the rotary knob and the
key, change to
the
<Flow Scheme
> screen.
OK
OK
OK
W16: Press 'OK' to return to menu selection !
Setting
Return flow
ml/min
30
[ 10 : 50 ]