Operating Manual
REINFUSION
Software 2.6
B|BRAUN
Chapter 6
Page 5 of 14
SHARING EXPERTISE
2.6_CA-EN_03
6.2 BLOOD REINFUSION
Stop the blood pump with the
key.
As long as the blood pump is running, the blood
reinfusion will not start !
The next steps are summarized in a Warning window.
Remove the arterial line from the patient’s arterial
access and connect the line to the 500 ml
compartment of the saline bag on the IV-pole.
Close the clamp of the reinfusion line.
Take the reinfusion line from the saline bag and
screw it to the port of the venous chamber.
Open the clamps of the reinfusion line and the
port.
Confirm the Warning window with the
key.
Start the blood pump with the
key
The default setting of the blood reinfusion volume is
300 ml.
OK
Start
Stop
Start
Stop
Warning
W21:
1) Connect art. line to saline solution bag
2) Connect reinfusion line to venous chamber
Press ‘OK’ to proceed!