Operating Manual
ALARMS AND PROBLEM CORRECTION
Software 2.6
B|BRAUN
Chapter 8
Page 20 of 27
SHARING EXPERTISE
2.6_CA-EN_03
CAUTION
Risk to patient due to impact on the patient’s fluid
balance.
Perform the balance reset only when you are
sure that the balancing error does not concern
the patient!
Balance Reset
Starting with a balance error > 200 g, the <
Balance
Reset
> menu item under <
Additional Functions
> can
be selected by turning the knob and pressing the
key. Warning W29: <Are you sure to reset Patient
Balance?> is displayed. Press
to proceed.
During a balance reset, the load cell is newly tared.
The data of the balance reset are saved and shown in
the Parameter Overview. Every reset performed in the
course of the therapy is saved and the values are
summated.
OK
W25: Balance error > 400 g