When a blood reinfusion volume of 150 ml has been
reached warning W41 appears:
open the clamp of the plasma line after the plasma
filter.
close the clamp on the venous line to the venous
chamber.
The saline solution is now pumped through the
membrane of the plasma filter to the plasma side of
the filter. In this manner, the plasma from the plasma
filter is also reinfused.
The blood pump stops automatically when the set
blood reinfusion volume is reached.
Remove the venous line from the patient’s venous
access. For the patient, the treatment is now
completed.
W41: Open plasma clamp and close venous clamp
W17 Blood reinfusion completed