Operating Manual
PREPARATION
Software 2.6
B|BRAUN
Chapter 3
Page 8 of 9
SHARING EXPERTISE
2.6_CA-EN_03
(1)
Connect the buffer line to the prepared saline
bag on the load cell.
(2)
Connect the ultrafiltrate lines to the three drain
bags.
(3)
Insert the buffer line into the holder provided on
the load cell.
Setting Up the Arterial Line
(1)
Place the arterial air chamber into the
holder.
(2)
Place the pump segment of the arterial line into
the blood pump.
(3)
Connect the arterial feeder line to the inlet of
the plasma filter.
(4)
Connect the arterial line to the prepared saline
bag which is hanging on the IV-pole.
(5)
Screw on the two pressure transducers as shown
in the Figure.
(6)
Fill a syringe (30 ml Omnifix® Perfusor syringe)
with heparin saline mixture and connect it with
the heparin line. Vent the heparin line manually
up to the T-piece. Make sure that no air bubbles
are left in the line. Mount the syringe on the
holder of the heparin pump.
Recommendation: 16 ml 0.9% NaCl + 4 ml
heparin (5000 IU/ml) corresponding to a
concentration of 1000 IU heparin/ml