Code
Alarm Text
Alarm Cause
Corrective Action
A 57
Air in dialysate line
DAD detects air
Ensure that:
The dialysate bags are full.
The clamps of the dialysate lines are open.
The seals of the dialysate bags are open.
The dialysate line is not damaged and the
connections to the bags are tight. Replace
the line if it is damaged.
A 58
Stop of blood pump too
long !
Blood pump stop > 120 s
Start the blood pump to eliminate the alarm
and to acknowledge the error.
A 59
Blood pump cover open Blood pump cover open,
magnetic sensor of pump
defective
Close the pump cover.
A 60
Plasma/ buffer pump
cover open
Plasma/buffer pump cover
open
Magnetic sensor of pump
defective
Close the pump cover.
A 61
Plasma return pump
cover open
Reinfusion pump cover
open
Magnetic sensor of pump
defective
Close the pump cover.
A 62
Dialysate pump cover
open
Cover of dialysate pump
open
Magnetic sensor of pump
defective
Close the pump cover.
A 63
Blood pump speed error Wrong speed of blood
pump
Defective blood pump
Ensure that:
The pump segment is correctly inserted in
the blood pump.
A 64
Plasma/buffer pump
speed error
Wrong speed of
plasma/buffer pumps
Pump defective
Ensure that:
The pump segment is correctly inserted in
the plasma/ buffer pump
A 65
Plasma return pump
speed error
Wrong speed of reinfusion
pump
Reinfusion pump
defective
Ensure that:
The pump segment is correctly inserted in
the reinfusion pump.
A 66
Dialysate pump speed
error
Wrong speed of dialysate
pump
Defective dialysate pump
Ensure that:
The pump segment is correctly inserted in
the dialysate pump.
A 67
Dialysate temperature
out of limits
Dialysate too warm (>
41.5
o
C for > 10 s)
Defective heating element
Close the cover of the plate warmer.
A 68
Excessive weight
change, check bags and
lines !
Weight variation
between 50 and 200 g for
more than 5 s or weight
variation > 200 g
Ensure that:
The bags are hanging motionless on the
load cell.
The lines are hanging free and do not pull
on the bags on the load cell.
The bags do not move too much.
This alarm is also activated if a bag has been